tuesday


tuesday went to SGH first for the commissioning of a Graseby syringe driver.
after that went to alexandra hospital to do preventive maintenance on a suction pump.
we went to clinic F to retrieve the machine and bring it back to Alexandra Hospital BME to do the testing as there is no place in the clinic to test.
while doing the battery test, we found out that the battery is totally flat so in the jobsheet must write change of battery needed.

last stop for tuesday was Raffles hospital.
they had some problems with the networking again.
this time the ALOKA guy, sam, came as well to configure the computer to the (ultrasound)ALOKA machine.

they realised that one of the cables was missing and therefore cannot transmit the data over to the other end so the sam has brought a cable and replace it and the problem is solved.
sometimes is not just any cables that looks alike can be connected to the system, we must read the manuals and see whats the internal wiring of the 16/32 pins cable are and the directions to connect them is very important.